RP-UPLC Analytical Method Development For Simultaneous Estimation Ritodrine By Analytical Qbd (Aqbd)) Approach
DOI:
https://doi.org/10.53555/AJBR.v27i4S.7414Keywords:
Ritodrine, QbD Approach, Method development, RP-UPLC, Desirability, ANOVAAbstract
A simple, Accurate, precise method was developed for the estimation of the Ritodrine in bulk and pharmaceutical dosage form. Chromatogram was run through Hibar100 x 2.1 mm, 2m. Mobile phase containing 0.1N OPA: Acetonitrile taken in the ratio 55(%v/v) and 45 was pumped through column at a flow rate of 0.9 ml/min. Temperature was maintained at 30°C. Optimized wavelength selected was ACQUITY TUV ChA 220.0 nm. Retention time of Ritodrine was found to be 1.065 min. %RSD of the Ritodrine was found to be 1.5. % Recovery was obtained as 99.95% for Ritodrine. Regression equation of Ritodrine is y = 69931x + 4571.9. Retention times were decreased and that run time was decreased, so the method developed was simple and economical that can be adopted in regular Quality control test in Industries.
Downloads
Published
Issue
Section
License
Copyright (c) 2024 Kasi Viswanadh Vuyyuri, Meenakshi Sharma (Author)